Product Description
📘 Product Overview
Field | Details |
---|---|
Manufacturer | Invensys (Foxboro, now Schneider Electric) |
Model Number | 873RS-AIYFNZ-7 |
Product Type | Electrochemical Analyzer for Resistivity Measurements |
Function | Measures resistivity of aqueous solutions with automatic temperature compensation |
System Compatibility | Compatible with 871CC sensors (0.1 cm⁻¹ cell factor); supports single or dual sensor setups |
Included Components | Analyzer unit, optional rear cover (BS805QK for plastic enclosure) |
⚙️ Hardware Specifications
Feature | Specification |
---|---|
Enclosure Type | Metal (NEMA 4X, CSA Enclosure 4X, IP-65), movable surface mount (Z option) |
Dimensions | 92(H) × 92(W) × 259(L) mm (3.6 × 3.6 × 10.1 in) |
Weight | 3.13 kg (6.9 lb) for movable surface mount |
Display | Four-digit red LED (14.2 mm height), visible up to 6 m (20 ft) |
Keypad | 8 keys (6 dual-function); includes Shift, Abso, Temp, Enter, etc. |
Sensor Inputs | Dual sensor inputs (Cell 1 and Cell 2) |
Memory | Non-volatile EEPROM for operating parameters and calibration data |
🔌 Electrical Specifications
Feature | Specification |
---|---|
Supply Voltage | 120 V AC (A option) |
Supply Frequency | 50 or 60 Hz, ±3 Hz |
Analog Outputs | Two isolated outputs: 4–20 mA (I option); Output #1 and #2 configurable |
Alarm Contacts | Two Form C contacts, 5 A noninductive, 125 V AC/30 V DC (min 1 A) |
Output Isolation | Isolated, powered outputs |
RFI Susceptibility | Compliant with European EMC Directive 89/336/EEC when cables are in grounded conduit |
📏 Measurement Specifications
Feature | Specification |
---|---|
Measurement Ranges | 0–2.000, 0–5.000, 0–10.00, 0–20.00 MΩ·cm |
Display Units | MΩ·cm or % (for ratio or % rejection) |
Temperature Range | Measurement: -17 to +150 °C (0 to 302 °F); Compensation: 0 to 120 °C (32 to 248 °F) |
Accuracy | ±0.5% of full-scale range at 25 °C |
Response Time | 3 s for 90% change (single sensor, no damping); 7 s (dual sensor) |
Damping | Configurable: 0, 10, 20, 40, 80, 160 s |
🔧 Configuration and Calibration
Feature | Details |
---|---|
Configuration Mode | Allows setting measurement range, alarms, diagnostics, and output characterization |
Security | User-configurable security code (default: 0800) |
Calibration | Electronic bench calibration; sensor-specific calibration with tCF and CF |
Sensor Compatibility | 871CC sensors (A2–F2: 100 kΩ thermistor; K2–M2: Pt 100 RTD) |
Cell Factor (CF) | Factory default: 0.1000 cm⁻¹; adjustable for specific sensor |
Temperature Compensation | Ultrapure water correction referenced to 25 °C or absolute (no compensation) |
🌡️ Environmental Specifications
Parameter | Specification |
---|---|
Operating Temperature | -25 to +55 °C (-13 to +131 °F) |
Storage Temperature | Not specified in document |
Relative Humidity | 5 to 95%, non-condensing |
Protection Rating | NEMA 4X, CSA Enclosure 4X, IEC IP-65 (metal enclosure) |
✅ Certifications
Standard | Details |
---|---|
FM | General purpose (FGZ); Nonincendive Class I, II, Div. 2, Groups A–D, F, G, T6 |
CSA (Canada) | Class I, Div. 2, Groups A–D, T6 (for 24 V, 100 V, 120 V AC) |
EMC | Complies with European EMC Directive 89/336/EEC (with proper cable conduit) |
RoHS/WEEE | Exempt from RoHS Directive; complies with WEEE Directive |
🛠️ Installation
Feature | Details |
---|---|
Mounting Options | Movable surface mount (Z); also supports panel, pipe, fixed surface (metal only) |
Wiring | Sensor, power, and alarm wires to TB1, TB2, TB3, TB4; separate conduits for sensor/output and power/alarms |
Conduit Requirements | 1/2-inch rigid metal or semi-rigid plastic; specific fittings (e.g., T&B #370) |
Grounding | Outer electrode (green wire) must be grounded for high-voltage solutions |
🚨 Alarms and Diagnostics
Feature | Details |
---|---|
Alarms | Two configurable alarms (HAC, LAC); set points adjustable 0 to full scale |
Alarm Timers | Trigger, feed, delay timers (0–99 min, 0–9 tenths/hundredths) |
Hysteresis | 0–99% of full scale, configurable |
Error Codes | Er 1 (instrument fault), Er 2 (temp error), Er 3 (range error), Er 4 (calibration error) |
Diagnostics | Troubleshooting for noisy signals, incorrect readings, gas bubbles, etc. |
📦 Ordering Information
Field | Details |
---|---|
Model Number | 873RS-AIYFNZ-7 |
Code Breakdown | A: 120 V AC; I: 4–20 mA outputs; Y: Pipe mount; F: Optional feature; NZ: Movable surface mount; -7: Additional options (not detailed in document) |
Accessories | Optional storm door, resistor kit for calibration, rear cover (plastic enclosure) |