Product Description
Product Overview
| Field | Details |
|---|---|
| Manufacturer | Honeywell Analytics |
| Part Number | 04230-A-1001 |
| Product Type | Handheld Interrogator SHC-1 |
| Function | Configures, aligns, calibrates, and diagnoses Searchline Excel, Optima, and Optima Plus gas detectors |
| Compatibility | Searchline Excel, Searchline Optima, Searchline Optima Plus |
| Associated Document | MAN0530 Issue 12 - 10/12 (Searchline Excel Technical Handbook) |
Functional Specifications
| Feature | Specification |
|---|---|
| Interface | Menu-style interface via RS485 serial link for commissioning, configuration, and diagnostics |
| Connection Method | Connects to receiver via DX/DVC100(M) series termination units or junction box with SHC Protection Device |
| Software Version | 4V0 |
| Key Functions | - System self-test - Alignment (Instrument Assisted Alignment, IAA) - Initialization - Time and date setting - Zero calibration - Force 4-20mA output - Return to normal 4-20mA output - Display calibration and configuration status - Display active faults and warnings - Soft reset |
| Alignment Process | - Displays SIG (current signal) and TGT (target signal) bar graphs - Ensures SIG ≥ TGT (minimum 70% of expected signal for path length) - Supports bayonet-style Trident and latch-style telescopes |
| Calibration | Inhibits 4-20mA output during zero calibration; requires no gas in beam path |
| Output Control | Adjustable 4-20mA output from 1 mA to 21 mA in 0.1 mA steps |
| Configuration Display | Shows gas, user, and installation parameters |
| Operating Modes | Toggles between Excel, Optima, and Optima Plus modes |
Physical Specifications
| Feature | Specification |
|---|---|
| Dimensions | Length: 190 mm (7.48 in) Width: 80 mm (3.15 in) Depth: 40 mm (1.57 in) |
| Weight | 0.5 kg (1.1 lb) |
| Material | Stainless Steel ANSI 316 |
Environmental Specifications
| Parameter | Specification |
|---|---|
| Operating Temperature | 0°C to +40°C (+32°F to +104°F); DMT evaluated: -25°C to +55°C (-13°F to +131°F) |
| Storage Temperature | -10°C to +40°C (+14°F to +104°F) |
| Operating Humidity | 0 to 99% RH (non-condensing) |
Electrical Specifications
| Feature | Specification |
|---|---|
| Power | Battery-powered; auto power-off after 5 minutes of non-use (disabled during configuration changes) |
| Fast Power-Off | Press ↓ and ESC simultaneously |
Regulatory Compliance
| Certification Type | Details |
|---|---|
| ATEX & IECEx | ATEX: Baseefa03ATEX0073 IECEx: BAS 09.0120 II 2G Ex ia IIC T4 Gb (Tamb -40°C to +40°C) |
| Standards | EN60079-0:2012, EN60079-11:2012 IEC60079-0:2011, IEC60079-11:2011 |
Ordering Information
| Field | Details |
|---|---|
| Part Number | 04230-A-1001: Handheld Interrogator ATEX Certified (4V0 software) for Optima, Optima Plus, and Excel |
| Related Parts | - 2104B2351: UL Certified Handheld Interrogator - 2104B2354: CSA/FM Certified Handheld Interrogator - 2108B1455: Handheld EEPROM upgrade (4V0) - 04320-A-1025: SHC Protection Device Module and Lead - 2104B6250: SHC Calibrator Connector System 10m |
| Accessories | Included in alignment kits |